multi EA 5000 Maintenance and care
165
} Cover the copper and brass wool with quartz wool.
} Carefully press the filled U-tube into the clamps.
} Connect the fast connector with hose 81 to the gas inlet limb with copper wool and
hose 82 to the gas outlet limb with brass wool.
} Check the system for leaks.
ü The detection module is ready for operation again.
14.14.3Regenerating the TIC reactor
WARNING
Risk of chemical burns
The TIC reactor is regenerated and cleaned with 40% phosphoric acid. Phosphoric acid
can irritate eyes, skin and mucous membranes.
¡ Wear protective clothing when handling the concentrated acid.
¡ Observe all notes and specifications in the safety data sheet.
NOTICE
Risk of leaks
A cannula that is too large will damage the septum on the septum port.
¡ Only use cannula with an outer diameter of 0.63mm for the septum port.
ð For TIC determination or TIC determination in differential mode: The TIC reactor
must be regenerated daily. How often this must be done is dependent on the TIC
content of the samples. For high TIC content, regenerate the TIC reactor more than
once daily.
ð Regeneration of the TIC reactor is also required after longer times of inactivity.
ð If only the TC or NPOC mode is used, regeneration of the TIC reactor is not required.
} Select the
System
|
Component test
menu.
} Select
Regeneration TIC reactor
from the list field in the
Device
tab.
} Click the
Regeneration TIC reactor
button.
} After prompt via software: Add 40% phosphoric acid to the TIC reactor via the sep-
tum port with the supplied 5ml syringe.
ü The TIC reactor is drained off and purged.
14.14.4Cleaning the TIC reactor
WARNING
Risk of chemical burns
The TIC reactor is regenerated and cleaned with 40% phosphoric acid. Phosphoric acid
can irritate eyes, skin and mucous membranes.
¡ Wear protective clothing when handling the concentrated acid.
¡ Observe all notes and specifications in the safety data sheet.

Table of Contents