Chlorine analysis with the Cl module multi EA 5000
60
8.2 Installation
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
} Place the detection module to the immediate left of the basic module.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module with the basic module via the interface:
CL-Coul interface on the rear of the detection module
CL-Coul interface on the rear of the basic module
} Connect the "waste" outlet with the exhaust hose. Connect the hose with the exhaust
system, or route the hose into a vent.