Maintenance and care multi EA 5000
158
} Place the lamp in the proper position when inserting: The pin on the
holder must fit into the groove of the lamp body.
} Fasten the new lamp with the 2 screws.
} Plug the connection plug back into the socket all the way.
} Fasten the side panel again.
ü The detection module is ready for operation again.
14.12.2Replacing the chemical ozone decomposer
Only for the MPO 5000
ð Replace the entire chemical ozone decomposer at least once annually.
ð Always replace the ozone decomposer when the smell of ozone is noticeable.
} Undo the hose from the chemical ozone decomposer. Do not pull the hose out of the
device!
} Pull the ozone decomposer out of the holding clamps.
} Press the new ozone decomposer into the holding clamps. Reattach the hose.
ü The detection module is ready for measurement again.