Troubleshooting multi EA 5000
110
The following status displays are possible:
Device not ready for measurement
Display Description
No sampling module dis-
played (display empty)
Device not ready for measurement, no sampling mod-
ule identified:
¡ Connect and activate at least one sampling mod-
ule
¡ Basic initialization
Auto-injector - syringe: ? No auto-injector syringe detected
¡ Initialize
¡ Insert auto-injector syringe and register it in mul-
tiWin ("Extras | New auto-injector syringe" menu
item)
13.4.4 Detectors
Example Cl-POT OK 3050
Meaning Detector status Current sensor value
All detector modules detected during initialization are displayed here. The following dis-
plays are possible:
Display Description
C-NDIR TC module or TOC module with NDIR detector
N-CLD CLD 5000 with chemiluminescence detector
S-UVFD UVFD 5000 and MPO 5000 with UV flourescence detector
S-Coul Coulometric Sulfur Module with microcoulometer
Cl-POT Cl module with "high sensitive" measuring cell
Cl-AMP smallCell Cl module with "sensitive" measuring cell
Cl-AMP largeCell Cl module with "high concentration" measuring cell
The respective status of the detector module is indicated by the color:
Display Description
Black Detector is active, status is queried and displayed (see example
above)
Gray Detector is inactive, status is not displayed
Green Detector OK, device ready for measurement (OK)
Or
Detector busy, measurement can only be started once the routine is
complete (detector-specific)
Red Error, see overview below