multi EA 5000 Troubleshooting
111
The following status displays are possible:
Device ready for measurement
Display Description
OK (green, black) Detector is ready for measurement
Device not ready for measurement - general
Display Description
No detector displayed (dis-
play empty)
No detector detected
¡ Activate detector
¡ Basic initialization
Communication error (red) Communication interrupted:
¡ Deactivate/activate the device
¡ Basic initialization
No connection (red) Connection interrupted:
¡ Check connection cable
¡ Deactivate/activate the device
¡ Basic initialization
Device not ready for measurement – Cl-POT
Display (red) Description
Inactive No cell detected:
¡ Insert cell
¡ Basic initialization
Drift exceeds range The current indicator drift is greater than the maxi-
mum positive drift set in the method or the max. neg-
ative drift defined in the system.
¡ Wait for the drift to go back into range (this is
normal immediately after cell maintenance or af-
ter commissioning the device)
¡ Check the maximum drift setting in the method
and increase as necessary (e.g., to 100Counts/
min).
¡ Access further values via the
System
|
Component
test
|
Chlorine
menu item.
End point routine required Indicator value outside the operating range of the
titration cell:
¡ Indicator value greater than 10000: End point
routine starts automatically
¡ Indicator value less than 1000: Start an end point
routine via the
System
|
End point routine
menu
item and follow the instructions
Cell temperature exceeds
range
Current cell temperature does not correspond to cell
temperature configured in the method:
¡ Wait until the desired cell temperature is
achieved
¡ Access the values via the
System
|
Component
test
|
Chlorine
menu item.
¡ Check the cell temperature settings for the
method and adjust as necessary
Device not ready for measurement – Cl-POT
Display (green) Description