14.11 Maintenance of the Cl module chlorine detector
14.11.1Replacing the sulfuric acid and cleaning the sulfuric acid container.
WARNING
Risk of chemical burns
Concentrated sulfuric acid can cause severe chemical burns!
¡ Before replacing the sulfuric acid: Switch off the gas supply via the software. Risk of
splashing if the gas supply is running.
¡ Wear protective clothing when working on the sulfuric acid container.
¡ Observe all instructions and specifications in the safety data sheet.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
The sulfuric acid absorbs the water that is produced during the combustion process. If
the acid content falls below 85% the sulfuric acid can no longer sufficiently dry the reac-
tion gas. Chlorine values that are too low are then measured.
1
2
3
4
5
6
7
3
5
4
Figure71Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment 2 Sulfuric acid container
3 Hose for measuring gas infeed 4 Connector
5 Banjo bolt 6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Analytik Jena multi EA 5000 and is the answer not in the manual?

Endress+Hauser Analytik Jena multi EA 5000 Specifications

General IconGeneral
BrandEndress+Hauser
ModelAnalytik Jena multi EA 5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals