Troubleshooting multi EA 5000
112
End point routine Automatic end point routine running:
¡ Wait for the end point routine to finish
Drift determination Drift determination immediately after titration or end
point routine:
¡ Wait until drift determination is complete (ap-
prox. 1minute)
Device not ready for measurement - CI-AMP
Display (red) Description
Inactive No cell detected:
¡ Insert cell
¡ Basic initialization
Device not ready for measurement - CI-AMP
Display (green) Description
Stand-by titration Interval titration running:
¡ Measurement start possible
End point routine End point routine running:
¡ Wait for the end point routine to finish
Device not ready for measurement – C-NDIR
Display (red) Description
Analog values warning Analog values outside range:
¡ Access the values via the
System
|
Component
test
|
NDIR
menu item.
Running-in time Detector not yet ready for operation:
¡ Wait for the run-in period after activation to com-
plete (approx. 30min)
Device not ready for measurement – S-UVFD or N-CLD
Display (red) Description
Preheating phase Detector not yet ready for operation:
¡ Wait for the run-in period after activation to com-
plete (approx. 30min)
Vacuum/pressure error Pressure in the detector outside the permitted range:
¡ See device faults of the N detector
13.5 Device faults on the basic module
Error Possible cause Remedy
Furnace does not heat Thermocouple connector
not connected
Connect the connector
(→ "Removing and in-
stalling the combustion
furnace" 144)
Temperature set incor-
rectly in the software
Check temperature con-
figuration in the method
No method loaded Load method
Malfunction in power
supply
Switch on the device
Check the internal fuses