multi EA 5000 Sulfur analysis with the Coulometric Sulfur Module
81
10.2 Installation
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡ Handle glass parts with extreme caution.
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡ Always connect the modules to power when they are switched off.
NOTICE
The detection module cannot be operated together with optical detectors.
¡ Do not connect the detection module together with an optical detector.
} Place the detection module to the left of the basic module.
} Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
} Connect the detection module with the basic module via the interface:
S-Coul interface on the rear of the detection module
S-Coul interface on the rear of the basic module
} Route the measuring gas hose (hose 71) from the basic module through the side
opening on the detection module. Connect the hose to the top of the NOx absorber.
} Connect the outlet of the HX absorber to hose 72.
Hose 72 is connected to the measuring cell later.
} Check that the NOx absorber and the HX absorber are connected via hose 73.
} Insert the electrodes into the measuring cell as displayed in the illustration.
} Insert the gas inlet tube into the measuring cell. Connect the gas inlet tube with hose
72 via the PTFE connector.
} Connect the electrodes to the "Generation" and "Indication" connections.
The connections cannot be mixed up.
ü Connection of the coulometric sulfur detector is complete.