Carbon analysis with the TOC module multi EA 5000
98
Figure63TOC detector, door opened
1 Water traps 2 Measuring gas hose from the basic
module (hose 80)
3 Condensate pump 4 TIC reactor
5 Cooling block (measuring gas dryer) 6 Halogen trap
7 Hose 81
Preparing the basic module
} Fill the TOC combustion tube as shown in the illustration.
} Insert the TOC combustion tube into the combustion furnace of the basic module.
Connect the oxygen supply (hose 3). Fasten the combustion tube with the tube
holder.
} Install the condensation coil in the basic module. Connect the condensation coil with
the TOC combustion tube with the spherical joint. Secure the spherical joint connec-
tion with a clamp.
} Connect the condensation coil to the TIC reactor in the detection module with hose
80. Route the hose through the side openings of the modules for this.