Troubleshooting multi EA 5000
126
Error Possible cause Remedy
Low results ¡ Catalyst depleted ¡ Exchange the catalyst
¡ System is leaking ¡ Check the sluice for leaks
¡ Replace septum
¡ Incorrect injection vol-
ume
¡ For manual sampling:
introduce the sample
volume set in the
method
¡ Phosphoric acid in the
TIC reactor depleted
¡ Regenerate the TIC reac-
tor, see (→ "Regenerat-
ing the TIC reactor"
165)
¡ Septum faulty ¡ Replace septum