Chlorine analysis with the Cl module multi EA 5000
68
The "sensitive" and "high con-
centration" measuring cells
Fill the measuring cells with fresh electrolyte solution daily:
¡ "sensitive" measuring cell: 15 to 20 ml
¡ "high concentration" measuring cell: 120 ml
The "high sensitive" measuring
cell
Fill the measuring cell daily with electrolyte solution: to approx. 65 ml
Change the electrolyte solution:
¡ Once per week
¡ If analytical problems occur
¡ If crystalline deposits form
Open the refill opening of the reference electrode during operation.