EasyManuals Logo

Endress+Hauser Analytik Jena multi EA 5000 User Manual

Endress+Hauser Analytik Jena multi EA 5000
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
multi EA 5000 Intended use
9
Parameter Description
TS Total sulfur
Organically bound sulfur is detected. Inorganic sulfur compounds are
only digested partially or not at all.
TN Total nitrogen
Organically bound nitrogen is detected. Inorganic nitrogen compounds
and molecular nitrogen (N
2
) are only digested partially or not at all.
TC Total carbon
The total organic carbon and elementary carbon and contained CO
2
is
detected. Inorganic carbon compounds are only digested partially or not
at all.
TX Total chlorine
The organically bound chlorine contained in the sample is detected.
Bromide and iodine compounds are only detected partially. Fluorine
compounds are not detected at all. The result is output as defined as to-
tal chlorine. Inorganic halogen compounds are only digested partially or
not at all.
EOX Extractable organically bound halogens
The EOX parameter is the sum of organically bound halogens (chlorine,
bromide, iodine) extracted under defined conditions from a water sam-
ple or from solids (sludge and sediment) using an organic solvent. Fluo-
rine is not detected.
AOX Adsorbable organically bound halogens
The AOX parameter is the sum of organically bound halogens (chlorine,
bromide, iodine) adsorbed under defined conditions (pH < 2 with
HNO
3
) from a water sample or a solid sample (sludge and sediment) in
active carbon using the vibration or column method. Fluorine is not de-
tected.
EC/OC Elementary carbon/organic carbon
The content of elementary carbon of a sample is detected directly after
the thermal desorption of the organic carbon. The amount of organic
carbon is then determined sequentially after thermal oxidation of the
remaining sample.
The following parameters can be determined from aqueous samples with the
multi EA 5000
:
Parameter Explanation
TC Total carbon
TIC Total inorganic carbon
The total inorganic carbon from carbonates, hydrocarbonates and dis-
solved carbon dioxide is determined.
TOC Total organic carbon
The total organic carbon content is determined mathematically by cal-
culating the difference: TOC = TC - TIC. Determination of the TC and TIC
content of the sample is required for this.
NPOC Non purgeable organic carbon
The total non-purgeable organic carbon content of a sample is de-
tected. Highly volatile organic compounds are not or only partially de-
tected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Analytik Jena multi EA 5000 and is the answer not in the manual?

Endress+Hauser Analytik Jena multi EA 5000 Specifications

General IconGeneral
BrandEndress+Hauser
ModelAnalytik Jena multi EA 5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals