Proline Promag 53 MODBUS RS485 Troubleshooting
Endress+Hauser 103
98 >< BATCH QUANT. 472 P: >< BATCH QUANTITY
$: # 472
Underbatching:
The minimum quantity was
not reached.
Overbatching:
The maximum permitted
batching quantity was
exceeded.
Underbatching:
1. Increase fixed correction quantity.
2. Valve closes too quickly with active after run
correction. Enter smaller after run as mean value.
3. If the batching quantity changes, the minimum
batching quantity must be adjusted.
Overbatching:
1. Reduce fixed correction quantity.
2. Valve closes too slowly with active after run
correction. Enter larger after run as mean value.
3. If the batching quantity changes, the maximum
batching quantity must be adjusted.
!
Note!
Please observe Note in error message No. 471
99 PROGRESS NOTE 473 P: PROGRESS NOTE
!: # 473
End of filling process
approaching.
The running filling process
has exceeded the predefined
batch quantity point for the
display warning message.
No measures required (if necessary prepare to replace
container).
122 > MAX.FLOW RATE 474 P: > MAX. FLOW RATE
$: # 474
Maximum flow value entered
is
exceeded.
Reduce the flow value.
!
Note!
Please observe Note in error message No. 471
MODBUS No. Device status message
(local display)
Cause Remedy / spare part
Register:
6859
Data type:
Integer
Register:
6821
Data type:
String (18 byte)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser MODBUS RS485 and is the answer not in the manual?

Endress+Hauser MODBUS RS485 Specifications

General IconGeneral
BrandEndress+Hauser
ModelMODBUS RS485
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals