EasyManua.ls Logo

Endress+Hauser MODBUS RS485 - System Error Messages

Endress+Hauser MODBUS RS485
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag 53 MODBUS RS485 Troubleshooting
Endress+Hauser 97
9.2 System error messages
Serious system errors are always recognized by the instrument as "Fault message", and are shown
as a lightning flash ($) on the display. Fault messages immediately affect the inputs and outputs.
Simulations and positive zero return, on the other hand, are classed and displayed as "Notice
messages".
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair.
Important procedures must be carried out before you return a flowmeter to Endress+Hauser.
ä 115
Always enclose a duly completed "Declaration of contamination" form. You will find a preprinted
blank of this form at the back of this manual.
!
Note!
The error types listed in the following correspond to the factory settings.
See the information on ä 60
MODBUS No. Device status message
(local display)
Cause Remedy (spare part ä 107)
Register:
6859
Data type:
Integer
Register:
6821
Data type:
String (18 byte)
Response to a fault message:
The value "NaN" (not a number) is
transmitted to the MODBUS master
instead of the current measured value.
Depicted on the local display:
S = System error
$ = Fault message (with an effect on the inputs and outputs)
! = Notice message (without any effect on the inputs and outputs)
1SYSTEM OK - There is no error present in the device
No. # 0xx Hardware error
2 CRITICAL FAIL. 001 S: CRITICAL FAILURE
$: # 001
Serious device error. Replace amplifier board.
3 AMP HW EEPROM 011 S: AMP HW EEPROM
$: # 011
Amplifier:
faulty EEPROM.
Replace amplifier board.
4AMP SW EEPROM012 S: AMP SW EEPROM
$: # 012
Amplifier:
Error accessing EEPROM data.
The EEPROM data blocks in which an error has
occurred are displayed in the
"TROUBLESHOOTING" function (No. 8047).
Press Enter to acknowledge the errors in question;
default values are automatically inserted instead of
the faulty parameter values.
!
Note!
The measuring device must be restarted after fault
elimination.
11 SENSOR HW DAT 031 S: SENSOR HW DAT
$: # 031
Sensor DAT:
1. S-DAT is defective.
2. S-DAT is not plugged into
the amplifier board or is
missing.
1. Check whether the S-DAT is correctly plugged
into the amplifier board.
2. Replace the S-DAT if it is defective.
Check that the new replacement DAT is
compatible with the measuring electronics.
Check the:
–Spare part set number
Hardware revision code
3. Replace measuring electronics boards if
necessary.
4. Plug the S-DAT into the amplifier board.
12 SENSOR SW DAT 032 S: SENSOR SW DAT
$: # 032
Sensor:
Error accessing the adjustment
values stored in the S-DAT.

Table of Contents

Related product manuals