Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 141
1. Call up the desired parameter.
2. On the right in the working area, mouse over the parameter.
A tool tip with remedy information for the diagnostic event appears.
12.4 Adapting the diagnostic information
12.4.1 Adapting the measured value and device status
Measured value status
Each diagnostic number is assigned a specific measured value status at the factory. The
user can change this assignment for specific diagnostic numbers via the Var. status xx
parameter.
Navigation path
"Expert" menu → System → Diagnostic handling → Condensed status →Assign behavior of
measured value status xx
1421-1
Var. status: 44
/ ../Condensed statu.
Var. status: 46
Dev.status: 44
Uncert.:Proc.(S)
A0019175-EN
Quality Substatus Status (hex) Event category Event class
GOOD None 0x80 – –
GOOD Maintenance required 0xA4 M Warning
GOOD Maintenance demanded 0xA8 M Warning
GOOD Function check 0xBC – –
BAD Maintenance alarm 0x24 F Alarm
BAD Process related, no maintenance 0x28 F Alarm
BAD Function check 0x3C C Warning
UNCERTAIN Maintenance demanded 0x68 M Warning
UNCERTAIN Process related, no maintenance 0x78 S Warning
Device status
Each diagnostic number is assigned a specific device status at the factory. The user can
change this assignment for specific diagnostic numbers via the Dev. status xx parameter.
Navigation path
"Expert" menu → System → Diagnostic handling → Condensed status →Assign behavior of
device status xx
1405-1
Inval.proc.cond
Var. status: 44
/ ../Condensed statu.
Var. status: 46
Dev.status: 44
A0019186-EN