Installation Proline Prowirl F 200 PROFIBUS PA
26 Endress+Hauser
6.2.4 Mounting the transmitter of the remote version
L
CAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
‣
Do not exceed the permitted maximum ambient temperature .
‣
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in
warm climatic regions.
L
CAUTION
Excessive force can damage the housing!
‣
Avoid excessive mechanical stress.
The transmitter of the remote version can be mounted in the following ways:
• Wall mounting
• Pipe mounting
Wall mounting
80 (3.15)
80 (3.15)
19 (0.6)
! 8.6 (0.39)
M 8
A0019864
8 Engineering unit mm (in)
1. Drill the holes.
2. Insert wall plugs into the drilled holes.
3. Screw in the securing screws slightly at first.
4. Fit the transmitter housing over the securing screws and mount in place.
5. Tighten the securing screws.