Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 171
Parameter Prerequisite Description User interface
Operating time from restart – Shows the time the device has been in
operation since the last device restart.
Days (d), hours (h),
minutes (m) and seconds
(s)
Operating time – Indicates how long the device has been
in operation.
Days (d), hours (h),
minutes (m) and seconds
(s)
12.7 Diagnostic list
In the Diagnostic list submenu, up to 5 currently pending diagnostic events can be
displayed along with the related diagnostic information. If more than 5 diagnostic events
are pending, the events with the highest priority are shown on the display.
Navigation path
Diagnostics menu → Diagnostic list submenu
Diagnostics
/../Diagnoselist
Diagnostics3
Diagnostics2
F273Mainelectronic
A0014006-EN
33 Taking the example of the local display
To call up the measures to rectify a diagnostic event:
• Via local display → 139
• Via "FieldCare" operating tool → 140
12.8 Event logbook
12.8.1 Event history
F
I1091Config.change
I1157Mem.err.ev.list
F311Electr.failure
/../Eventlist
0d01h19m10s
A0014008-EN
34 Taking the example of the local display
To call up the measures to rectify a diagnostic event:
• Via local display → 139
• Via "FieldCare" operating tool → 140
For filtering the displayed event messages → 171
12.8.2 Filtering the event logbook
Using the Filter options parameter, you can define which category of event messages is
displayed in the Events list submenu.
Navigation path
"Diagnostics" menu → Event logbook → Filter options