EasyManua.ls Logo

Endress+Hauser Proline 200 - Operating Philosophy

Endress+Hauser Proline 200
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prowirl F 200 PROFIBUS PA Operation options
Endress+Hauser 47
8.2.2 Operating philosophy
The individual parts of the operating menu are assigned to certain user roles (operator,
maintenance etc.). Each user role contains typical tasks within the device lifecycle.
Menu/parameter User role and tasks Content/meaning
Language task-oriented Role "Operator", "Maintenance"
Tasks during operation:
Configuring the operational display
Reading measured values
Defining the operating language
Operation Configuring the operational display (e.g. display format,
display contrast)
Resetting and controlling totalizers
Setup "Maintenance" role
Commissioning:
Configuration of the measurement
Configuration of the inputs and
outputs
Wizards for fast commissioning:
Configure the outputs
Configuring the operational display
Define the output conditioning
Set the low flow cut off
Advanced setup
For more customized configuration of the measurement
(adaptation to special measuring conditions)
Configuration of totalizers
Administration (define access code, reset measuring device)
Diagnostics "Maintenance" role
Fault elimination:
Diagnostics and elimination of
process and device errors
Measured value simulation
Contains all parameters for error detection and analyzing
process and device errors:
Diagnostic list
Contains up to 5 currently pending diagnostic messages.
Event logbook
Contains up to 20 or 100 (order option " Extended
HistoROM") event messages that have occurred.
Device information
Contains information for identifying the device.
Measured values
Contains all current measured values.
Analog inputs
Is used to display the analog input.
Data logging
(Order option "Extended HistoROM")
Storage and visualization of up to 1000 measured values
Heartbeat
The functionality of the device is checked on demand and the
verification results are documented.
Simulation
Is used to simulate measured values or output values.
Expert function-oriented Tasks that require detailed knowledge
of the function of the device:
Commissioning measurements under
difficult conditions
Optimal adaptation of the
measurement to difficult conditions
Detailed configuration of the
communication interface
Error diagnostics in difficult cases
Contains all the parameters of the device and makes it possible
to access these parameters directly using an access code. The
structure of this menu is based on the function blocks of the
device:
System
Contains all higher-order device parameters which do not
concern the measurement or the communication interface.
Sensor
Configuration of the measurement.
Output
Configuration of the pulse/frequency/status output.
Communication
Configuration of the digital communication interface.
Submenus for function blocks (e.g. "Analog Inputs")
Configuration of function blocks.
Application
Configuration of the functions that go beyond the actual
measurement (e.g. totalizer).
Diagnostics
Error detection and analysis of process and device errors and
for device simulation and Heartbeat Technology.

Table of Contents

Other manuals for Endress+Hauser Proline 200

Related product manuals