Parameter Prerequisite Description User interface
Compressibility factor The following conditions are met:
Order code for "Sensor version", option
"Mass flow"
In the Select medium parameter, the
Gas option or Steam option is selected.
Displays the compressibility factor
currently calculated.
0 to 2
Degrees of superheat In the Select medium parameter, the
Steam option is selected.
Displays the degree of superheating
currently calculated.
0 to 500 K
11.4.2 Totalizer
The Totalizer submenu contains all the parameters needed to display the current
measured values for every totalizer.
Navigation
"Diagnostics" menu → Measured values → Totalizer 1 to 3
Totalizer 1 to 3
Assign process variable
Totalizer value 1 to 3
Totalizer status 1 to 3
Totalizer status (Hex) 1 to 3
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Assign process variable Select process variable for
totalizer.
Volume flow
Mass flow
Corrected volume
flow
Total mass flow
*
Condensate mass
flow
*
Energy flow
*
Heat flow
difference
*
Totalizer 1:
Volume flow
Totalizer 2: Mass
flow
Totalizer 3:
Corrected volume
flow
Totalizer value 1 to 3 In the Assign process variable
parameter one of the following
options is selected:
Volume flow
Mass flow
Corrected volume flow
Total mass flow
Condensate mass flow
Energy flow
Heat flow difference
Displays the current totalizer
counter value.
Signed floating-point
number
0 m³