Proline Prowirl F 200 PROFIBUS PA Technical data
Endress+Hauser 199
Device plugs
IP67, only in screwed situation
Vibration resistance For compact/remote version made of coated aluminum and remote version made of
stainless steel:
Acceleration up to 2 g (if gain set to factory setting), 10 to 500 Hz, following IEC
60068-2-6
For the compact version made of stainless steel:
Acceleration up to 1 g (if gain set to factory setting), 10 to 500 Hz, following IEC
60068-2-6
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
For details, refer to the Declaration of Conformity.
16.9 Process
Medium temperature range
DSC sensor
3)
Order code for "Sensor version":
Option 1 "Volume flow, basis":
–40 to +260 °C (–40 to +500 °F), stainless steel
Option 2 "Volume flow, high-temperature/low-temperature":
–200 to +400 °C (–328 to +752 °F), stainless steel
Option 3 "Mass flow (integrated temperature measurement)":
–200 to +400 °C (–328 to +752 °F), stainless steel
Order code for "Sensor option":
Option CD "Harsh environment
4)
, DSC sensor components Alloy C22":
–200 to +400 °C (–328 to +752 °F), DSC sensor Alloy C22
Option CE "Harsh process
5)
, wetted parts Alloy C22, (including option CD)":
–40 to +260 °C (–40 to +500 °F), sensor and DSC sensor Alloy C22
Seals
–200 to +400 °C (–328 to +752 °F) for graphite (standard)
–15 to +175 °C (+5 to +347 °F) for Viton
–20 to +275 °C (–4 to +527 °F) for Kalrez
–200 to +260 °C (–328 to +500 °F) for Gylon
Pressure-temperature
ratings
An overview of the pressure-temperature ratings for the process connections is
provided in the "Technical Information" document
Secondary containment
pressure rating
The following overpressure resistance values apply to the sensor shaft in the event of a
membrane rupture:
Sensor version Overpressure, sensor shaft
in [bar a]
Volume flow, basis 200
Volume flow, high-temperature/low-temperature 200
Mass flow (integrated temperature measurement) 200
3) Capacitance sensor
4) Aggressive atmosphere (salts or chloride in the air)
5) Aggressive medium (risk of corrosion due to chloride, for example)