EasyManua.ls Logo

Endress+Hauser PROline promag 53 - Page 49

Endress+Hauser PROline promag 53
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PROline Promag 53 4 Wiring
Endress+Hauser 49
Cable termination for the remote version
Promag W / Promag P
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Fit the fine-wire cores with cable end sleeves (Detail B).
" Caution!
When fitting the connectors, pay attention to the following points:
Signal cable
Make sure that the cable end sleeves do not touch the wire shield on the sensor side.
Minimum distance = 1 mm (exception “GND” = green cable).
Coil current cable
Insulate one core of the three-core wire at the level of the core reinforcement; you only require two cores for the
connection.
TRANSMITTER
Signal cable Coil current cable
SENSOR
Signal cable Coil current cable
F06-5xxxxxxx-04-11-08-en-000
Dimensions in inches (mm)
F06-5xxxxxxx-04-11-08-xx-000
Dimensions in inches (mm)
F06-5xFxxxxx-04-11-08-ae-000
Dimensions in inches (mm)
F06-5xFxxxxx-04-11-08-xx-000ae
Dimensions in inches (mm)

Table of Contents

Other manuals for Endress+Hauser PROline promag 53

Related product manuals