Proline Promag L 400 PROFIBUS DP Installation
Endress+Hauser 31
Step Fixing screw Tightening torques for housing made of:
Aluminum Plastic
1 Housing cover 2.5 Nm (1.8 lbf ft) 1 Nm (0.7 lbf ft)
5 Smart sensor electronics module 0.6 Nm (0.4 lbf ft)
7 Main electronics module 1.5 Nm (1.1 lbf ft)
10 Transmitter housing 5.5 Nm (4.1 lbf ft)
NOTICE
Plug of the smart sensor electronics module connected incorrectly!
No measuring signal is output.
Plug in the plug of the smart sensor electronics module as per the coding.
A0021585
Reverse the procedure to reassemble the measuring device.
6.2.6 Turning the display module
The display module can be turned to optimize display readability and operability.
TX 20
4 x
1.
2.
PUSH TO
REMOVE
PUSH TO
REMOVE
3.
4.
A0021617
1. Loosen the fixing screws of the housing cover (when reassembling, pay attention to
the tightening torque (→  32)).
2. Open the housing cover.
3. Unlock the display module.
4. Pull out the display module and turn it to the desired position in increments of 90°.