Proline Promag L 400 PROFIBUS DP Electrical connection
Endress+Hauser 35
Grounding is by means of the ground terminal provided for the purpose inside the
connection housing. The stripped and twisted lengths of cable shield to the ground
terminal must be as short as possible.
Cable diameter
• Cable glands supplied:
– For standard cable: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
– For reinforced cable: M20 × 1.5 with cable 9.5 to 16 mm (0.37 to 0.63 in)
• (Plug-in) spring terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
7.1.3 Terminal assignment
Transmitter
PROFIBUS DP connection version
The sensor can be ordered with terminals or a device plug.
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Terminals Terminals • Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Device plug Terminals • Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
A0020426
1 Supply voltage (wide range power unit)
2 PROFIBUS DP
Supply voltage
Order code for "Power supply" Terminal numbers
1 (L+/L) 2 (L-/N)
Option L
(wide range power unit)
AC100 to 240 V
AC/DC24 V