Proline Promag L 400 PROFIBUS DP Commissioning
Endress+Hauser 83
10.6.7 Configuring empty pipe detection
The Empty pipe detection wizard guides you systematically through all the parameters
that have to be set for configuring the low flow cut off.
Navigation
"Setup" menu → Empty pipe detection
Structure of the wizard
Empty pipe det.
Response time
End of wizard
New adjustment
Cancel
Empty pipe adj.
Progress
Full pipe adjust.
Ok
Busy
Not ok
On
Switch point EPD
Off
A0017210-EN
28 "Empty pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameter Description Selection / User interface /
User entry
Factory setting
Empty pipe detection Switch empty pipe detection on and off. • Off
• On
Off
New adjustment Select type of adjustment. • Cancel
• Empty pipe adjust
• Full pipe adjust
Cancel
Progress • Ok
• Busy
• Not ok
Switch point empty pipe detection Enter hysteresis in %, below this value the
measuring tube will detected as empty.
0 to 100 % 50 %
Response time empty pipe detection Enter the time before diagnostic message
S862 ''Pipe empty'' is displayed for empty
pipe detection.
0 to 100 s 1 s