EasyManua.ls Logo

EXFO MAX-890 - Page 148

EXFO MAX-890
495 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test Setup - Test Configurator, Timer, and System
140 MAX-800 Series
RFC 2544 - Global
Coupled is available with Bidirectional flow direction. When the
Coupled check box is:
Cleared, the decision to increase or decrease the TX Rate of the
next iteration is taken independently per direction.
Selected, the decision to increase or decrease the TX Rate of the
next iteration is taken commonly for both directions.
Rate Unit determines the unit used to display the rate values: %,
Mbit/s, or Gbit/s.
Pass/Fail Verdict check box when selected (default) enables the use
of the pass/fail verdict.
Subtests and Estimated Time
Subtests allows enabling the Throughput, Back-to-Back, Frame Loss,
and Latency subtests individually.
Estimated Time (H:MM) indicates the estimated time required to
complete each subtest at best conditions. The total estimated time to
complete all subtests is also displayed.
Frame Distribution
Frame Distribution allows selecting either RFC 2544 (default) or User
Defined distribution.
Quantity is only available when User Defined is selected and allows
selecting the number of frames, from 1 to 10 (7 by default), in the
distribution.

Table of Contents