Power Failure Recovery
404 MAX-800 Series
Enabling Power Failure Recovery
Enabling Power Failure Recovery
To enable the automatic power failure recovery:
1. Enable launching the application when starting the MAX-800 Series
(refer to the MAX-800 Series user guide for more information):
From Mini ToolBox, tap on the System Settings button, the Startup
Applications button, and select the model’s check box.
2. Enable the MAX-800 Series automatic power on feature (refer to the
MAX-800 Series user guide for more information):
2a. From Mini ToolBox, tap on the System Settings button, and the
Startup Applications button.
2b. Select the Power on the unit when AC outlet is connected or
after power outage check box.
3. Make sure that Windows does not require a user name and password.
The MAX-800 Series is set to require user name and password by
default. To disable Windows user name and password:
3a. From Mini ToolBox, tap on the System Settings button and the
Automatic Logon button.
3b. Clear the User must enter a user name and password to use
this computer check box and enter the password to confirm.
Note: The power failure recovery is not used when the application closes
normally.