Test Setup - Test Configurator, Timer, and System
MaxTester 189
Streams - Profile
Shaping button
For Burst and n-Burst TX modes.
Burst Duty Cycle (%) represents the burst duration within the
burst period: 1 to 100 percent (default is 50 percent).
Period represents the burst pattern duration: 1 to 8000
milliseconds (default is 1000 ms).
Unit choices are ms (default) and s.
Burst Count, available with n-Burst TX Mode, represents the
number of times the burst will be repeated: 1 (default) to 255.
For Ramp and n-Ramp TX modes.
Ramp Nb. of Steps represents the number of steps within the
ramp: 2 to 100 (default is 10).
Step Time represents the duration of each step: 100 to 8000
milliseconds (default is 1000 ms).
Unit choices are ms (default) and s.
Ramp Cycle Count, available with n-Ramp TX mode only,
represents the number of times the ramp will be repeated:
1 (default) to 255.