Test Control
MaxTester 399
Save/Load Button
Save/Load Tab
Tap the Save/Load button and the Save/Load tab.
The save function stores the configuration of the unit including all test
settings to a file.
To save a configuration:
1. Select the file location (see File location on page 398).
2. Tap on the Save button.
3. Type the name of the configuration file to be saved and a description
(Config Summary) if needed.
4. Select the Add to Favorites check box to save the configuration file in
the Favorites list.
5. Tap OK.
The load function opens and applies the test configuration from a
previously saved configuration file.
To load a configuration:
1. Select the file location (see File location on page 398).
2. Select the file from the list.
3. Select or clear the Overwrite report settings check box as required.
The Overwrite report settings check box when selected (default)
replaces the current report settings by those from the configuration
that is loaded.
4. Tap the Load button.
Note: Configuration file has a limited backward compatibility. (Typically the
backward compatibility period is one year or three software releases.)