Test Setup - Test Configurator, Timer, and System
84 MAX-800 Series
BERT and Unframed BERT
Multi-Pattern Configuration
Note: Only available for DSN/PDH BERT test application with DS1 signal (DS0
disabled) and when the pattern is Multi-Pattern.
This feature allows sending each pattern for a specific duration,
sequentially and continuously.
Pattern represents a sequence of patterns that will be generated: 1111,
1in8, 2in8, 3in24, and QRSS.
Enable allows enabling the generation of each pattern individually in
the pattern sequence. All patterns are enabled by default. All 1’s (1111)
is disabled for Unframed test.
Individual Pattern Duration specifies the transmission duration of
each pattern: 15 s, 30 s, 45 s, 1min, 2min, 3min (default),... up to
15 min.
Restore < Test Application > Defaults
Reverts the current test application to its default factory settings.