Chapter 8: Diagnostic and Repair Procedures
8-23
The rejector does not function.
1. Is the rejector activated? Turn the Rejector Status function on.
2. If you hear a clicking sound, but the rejector does not activate,
the air pressure is not turned on. Turn on air pressure to the
rejector.
3. The rejector Pulse Width function may need adjustment.
Contact IDC Customer Support for assistance.
4. Disconnect the wire at TB1-26 and use a digital voltmeter to
measure the resistance between the wire and TB1-25. The resis-
tance should be 165 ohms (±10%).
5. Turn the FT-50 on and off and then use the Fire Rejector On
Demand function to fire the rejector. LED 2 on the processor
card should flash for each reject pulse.
• If the LED flashes, but the rejector does not fire; replace the
power supply.
• If the LED does not flash, replace the processor card.
Unstable Main Power
Surges, sags, and transients in the AC mains can cause the
inspection system to falsely reject good containers. If this occurs,
try the following procedures.
1. Check the AC power mains powering the inspection unit using
line voltage measuring equipment to detect surges, sags, and
transients.
2. Check the AC power mains neutral and ground lines for
common-mode voltages and currents.
Industrial Dynamics Company, Ltd. Document 40255-0600