EM100-F User Manual
0084293
Page 29 of 93
5.2. Calibration Commands – CE, CM’n’, CI, MR, DS, DP, CZ, CG, ZT, FD,
IZ, ZR, ZI, TM, TN, ZN, AG, AZ, WT
Note: All changes to the calibration commands must be stored in non-volatile memory using the
‘CS’ command.
5.2.1. CE Read TAC Counter/Open Calibration Sequence
With this command, you can read the TAC counter (Traceable Access Code) or you can open a
calibration sequence.
Calibration Sequence Active.
This command must be issued PRIOR to any attempt to set the parameters in the calibration group
of commands. In ‘legal-for-trade’ applications the TAC counter can be used to check if critical
parameters have been changed without re-verification. After each calibration save (‘CS’) the TAC
counter increases by 1.
5.2.2. CM ‘n’ Set Maximum Output Value
This command (‘CM n’ with n=1, 2 or 3) is used to set up the maximum output value (respective the
switching point in multi range applications). Permitted values are from 0 to 99999.
Request: TAC Counter CE11.
Calibration Sequence Active.
This value will determine the point at which the output will change to ‘oooooooo’, signifying over-
range respective the point at which the output will change the measuring range/interval size.
CM 2=0 (means CM 2
Not used)
CM 3=0 (means CM 3
Not used)
Interval
Not used)
Interval