3.5. Calibration
3.5.1. Calibration Commands
The calibration of EM100 is only possible after opening a calibration sequence. See ‘Calibration
Commands’ Chapter for more details on the following commands.
Command CE: Calibration Enable – Returns the current TAC value.
Command CM1/CM2/CM3: Calibrate Maximum Display – Sets the max. allowable display value.
Command CI: Calibrate Minimum – Sets the minimum allowable display value.
Command MR: Multi-Range/Multi-Interval.
Command DS: Display Step Size – Sets the output incremental step size.
Command DP: Decimal Point – Sets the position of the output decimal point.
Command CZ: Calibrate Zero – Sets the system zero-point.
Command CG: Calibrate Gain – Sets the system gain.
Command ZT: Zero Tracking – Enabled/disabled.
Command IZ: If applicable: Correction of System Zero.
Command TM: If applicable: Tare Mode (check if ‘legal-for-trade’ application).
Command ZR: If applicable: Zero Range – sets the zero range manually.
Command ZI: If applicable: Initial Zero Range.
Command WT: If applicable: Warm-Up Time.
Command FD: If applicable: Reset to factory default settings.
Command CS: Save Calibration data (TAC automatically incremented by 1).
3.5.2. Preparing for Calibration
• Check the Maximum value of the display is set sufficiently for the application (see ‘CM n’).
• Check the Minimum value of the display is set sufficiently for the application (see ‘CI’).
• Check the Zero-Tracking has been turned off (see ‘ZT’).
• Check the Warm-Up time is not set to a high value (see ‘WT’).
• Check the No-Motion window conditions reasonably defined (Default: NR=1, NT=1000).