EM100-F User Manual
0084293
Page 30 of 93
It is necessary: 1 ≤ MAX1 < MAX2 < MAX3 ≤ 99999.
Note: The range, in which a scale can be set to zero (‘SZ’) or automatic zero-tracking (‘ZT’) is
active, is ±2% of CM value.
Factory Default: CM 1=99999, CM 2=0, CM 3=0.
5.2.3. CI Set Minimum Output Value
This command is used to set up the minimum output value. Permitted values are –99999 to 0. This
value will determine the point at which the output will change to ‘uuuuuuuu’, signifying ‘under-
range’ condition.
Request: TAC Counter CE11
Calibration Sequence Active
Note: In bipolar applications (e.g. force- or torque measurements) this parameter defines the max.
output value for input signals with negative sign.
Factory Default: –99999.
5.2.4. MR Set Multi-Range/Multi-interval
This command is only relevant, if CM 2 > 0 or CM 3 > 0. Is this the case, then this command
defines, if the application is multi-range or multi-interval. Permitted values are 0 (Multi-interval) or 1
(Multi-range).
Request: MR=0 (Multi-Interval).
Request: TAC Counter CE11.
Calibration Sequence Active.
Setup: MR=1 (Multi-Range).
Note: Single range applications ignore this parameter.
Factory Default: 0 [Single Range].