5.2.12. ZR Zero Range
Sets the zero range manually this is the range in increments within which the weighing scale can
be zeroed. Issuing the ‘ZR’ command without any parameter will return the current value.
Permitted values are between the lower limit of 0 and the upper limit of 999999. A value of zero
enables the standard zero range of ±2% of max.
Master Sends
Slave Responds With
Meaning
CE
E+00011 (example)
Request: TAC Counter CE11.
CE 11
OK
Calibration Sequence Active.
ZR 100
OK
Setup: Zero Range=100dec.
Factory Default: 0.
5.2.13. ZI Initial Zero Range
Define the initial zero range (0...999999dec). If ‘ZI’ is non-zero the device will perform an automatic
Set-Zero when the weight stabilizes with the No-Motion settings, the weight is within the ZIrange
and warm-up time has expired.
Master Sends
Slave Responds With
Meaning
CE
E+00011 (example)
Request: TAC Counter CE11.
CE 11
OK
Calibration Sequence Active.
ZI 100
OK
Setup: Initial Zero Range=100dec.
Factory Default: 0.
5.2.14. TM Tare Mode
This command sets the tare mode. The tare modes are defined in the table below.
Master Sends
Slave Responds With
Meaning
CE
E+00011 (example)
Request: TAC Counter CE11.
CE 11
OK
Calibration Sequence Active.
TM 1
OK
Setup: Tare Mode=1.
TM
Allow Tare of
Negative Values
Clear Preset Tare at
Return to Range 1
0 (Default)
Yes
Yes
1
No
Yes
2
Yes
No
3
No
No
For OIML R76 compatible applications a tare mode of 1 must be used.
Factory Default: 0.