5.4.3. UR Update Rate & Averaging
Depending on the selected filter mode this command defines an averaging for the output value. The
permitted settings are from 0 to 7 (see table below). The average value is always calculated from
2
UR
measurement values.
Master Sends
Slave Responds With
Meaning
UR
U:00003
Request: Averaging of 8 Values.
UR 7
OK
Setup: Averaging of 128 Values.
WP
OK
Setup Saved.
UR
0
1
2
3
4
5
6
7
Average of 2
UR
Values
1
2
4
8
16
32
64
128
Factory Default: 0 [No Averaging].