EM100-F User Manual
0084293
Page 31 of 93
5.2.5. DS Set Display Step Size
This command allows the output to step up or down by a unit other than 1. Permitted values are 1,
2, 5, 10, 20, 50, 100, 200 and 500.
Request: TAC Counter CE11.
Calibration Sequence Active.
‘Legal-for-trade’ applications allow for up to 10,000 intervals. The allowed step size must be
considered.
Factory Default: 1.
5.2.6. DP Set Decimal Point Position
This command allows the decimal point to be positioned anywhere between left-most and right-most
digits of the 5-digit output result. Permitted values are 0, 1, 2, 3, 4 and 5. Position 0 means no
decimal point.
Request: Position of Decimal Point.
Request: TAC Counter CE11.
Calibration Sequence Active.
Factory default: 3.
5.2.7. CZ Set Calibration Zero Point
This is the reference point for all weight calculations and is subject to TAC control.
Request: TAC Counter CE11.
Calibration Sequence Active.
Factory Default: Approx. 0mV/V input signal.