EM100-F User Manual
0084293
Page 68 of 93
5.13.13. PD8 In-Flight Delay After Fine Valve Shut-Off
After the fine valve shuts off, a weight peak may occur because of a surge or splash of the product
being filled.
Request: Delay Time=100ms.
Setup: Set Delay Time to 250ms.
Range: 0-65535ms.
5.13.14. PD9 Dosed Weight Average Time
This is the time the filled weight average is calculated. This function allows you to acquire the
correct filled weight despite possible vibrations on a (rotating) filling machine.
Request: Average Time=100ms.
Setup: Set Average Time to 500ms.
Range: 0-65535ms.
5.13.15. PD10 Zero Tolerance
Load-cell zero tolerance check (in increments). This is the allowable deviation from the unloaded
load cell zero (no bottle or box). If the unloaded load-cell zero is outside this window, the filling
process will not start. This function can easily check to see if a load-cell has been over-loaded.
Request: Load-Cell Zero Check Tol=100d.
Setup: Set Tolerance to 5,000d.
Range: 0…99999 increments.
Example: PD10 5000 means that the zero point of unloaded load-cell must be within the window
±5,000.0g relative to the reference zero [when cal. 1 kg=10,000 increments].
5.13.16. PD11 Tare Reference Weight
Tare reference weight (in increments). This is the nominal weight of the empty bottle/box.
Request: Tare Reference Weight=400d.
Setup: Set Tare Reference Weight=230d.
Range: 0…99999 increments.