EM100-C User Manual
0084292
Page 23 of 81
4. Commands Overview
Command Factory Default Parameter Range Description
A’n’ Zero. 0…2dec.
Allocation Source for a Set-Point
(Gross, Net or Average).
Set Device Address (0=perm active).
Absolute Gain Calibration.
Absolute Zero-Point Calibration.
BR 9600. 9600…460800dec. Baud Rate.
CE No Default. 0…65535dec.
Open Calibration Sequence. Read
TAC Counter.
CG 20000. 1…99999dec.
Set Calibration Gain (Span) at Load
> Zero.
Close Device Communication.
CM1 99999. 1…99999dec. Set Maximum Output Value (n=1).
Set Maximum Output Value (n=2).
Set Maximum Output Value (n=3).
CTR Zero. 0…1dec. CAN Termination Resistor Selector.
Set Calibration Zero Point (no-load).
Set Decimal Point Position.
DS 1.
100, 200, 500dec.
Set Display Step Size.
DT 50. 0…65535dec.
Calculation Time for Short-Time
Average (ms).
FD No Default. N/A Restore Factory Default Settings.
Identify Firmware Version.
Filter Cut-Off Frequency.
FPN No Default. N/A Get Device ID.
Get Triggered Average Value.
GL No Default. N/A
Get Data String (Average, Gross,
Status).
GM No Default. N/A Get Peak Value.