ProSim 8
Users Manual
48
Pressure, heart rate, pulse volume, brand, and wave are set through the front-
panel controls to simulate different patient conditions. Arrhythmia waveforms are
also simulated in the NIBP simulation (set through ECG). To set the Blood
Pressure:
1. Push or to highlight the Pressure variable.
2. Push .
Systolic and diastolic pressures are set separately.
3. Push or to highlight the systolic or diastolic pressure value.
4. Push or to increase or decrease the pressure value in 1 mmHg steps.
Note
When you hold down the direction key, the step size changes to
10 mmHg until the key is released.
5. Push to set the pressure and go back to the NIBP screen.
Note
Heart Rate may not show in the NIBP screen if an arrhythmia or
performance waveform is set.
To set the heart rate:
1. Push or to highlight the Heart Rate variable.
2. Push .
3. Push or to increase or decrease the heart rate in 1 BPM steps. The
range is 30 BPM to 240 BPM.
Note
When you hold down the direction key, the step size changes to
10 BPM until the key is released.
4. Push to set the heart rate and go back to the NIBP screen.
Note
Arrhythmia may show on the display instead of Heart Rate when
an arrhythmia or performance waveform is set under the ECG
function. The Mono VTach waveform is the one exception. You can
only change the heart beat to normal sinus rhythm in the NIBP
screen.
To set the heart beat waveform to normal sinus rhythm:
1. Push or to highlight the Arrhythmia variable.
2. Push .
3. Push or to increase or decrease the heart rate.
4. Push to set the heart rate and go back to the NIBP screen.