5500A
Service Manual
6-8
6-6. Leveled Sine Wave Specifications
Table 6-3. Leveled Sine Wave Specifications
Leveled Sine Wave
Frequency Range
Characteristics
into 50 Ω
50 kHz
(reference)
50 kHz to
100 MHz
100 MHz to
300 MHz
300 MHz to
600 MHz
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p) 5 mV to 5.5 V
Resolution < 100 mV:3 digits
≥ 100 mV:4 digits
Adjustment Range continuously adjustable
1-Year Absolute
Uncertainty,
tcal ± 5 °C
± (2% of
output
+ 300 µV)
± (3.5% of
output
+ 300 µV)
± (4% of output
+ 300 µV)
± (6% of output
+ 300 µV)
Flatness (relative to
50 kHz)
not applicable ± (1.5% of
output
+ 100 µV)
± (2% of output
+ 100 µV)
± (4% of output
+ 100 µV)
Short-Term Amplitude
Stability
≤ 1% [1]
Frequency Characteristics
Resolution 10 kHz
1-Year Absolute
Uncertainty,
tcal ± 5 °C
± 2.5 ppm
Distortion Characteristics
2nd Harmonic ≤ -33 dBc
3rd and Higher
Harmonics
≤ -38 dBc
[1] Within one hour after reference amplitude setting, provided temperature varies no more than ± 5°C.