Introduction and Specifications
Specifications
1
1-15
1-10. Capacitance Specifications
Ranges
Absolute Uncertainty, tcal ± 5 °C
± (% of output + nF)
Resolution
Frequency
90 days 1 year
Allowed Typical for
<1% Error
0.33 to 0.4999 nF 0.38% 0.01 nF 0.5% 0.01 nF 0.1 pF 50 to 1000 Hz 10 kHz
0.5 to 1.0999 nF 0.38 0.01 0.5 0.01 0.1 pF 50 to 1000 Hz 10 kHz
1.1 to 3.2999 nF 0.38 0.01 0.5 0.01 0.1 pF 50 to 1000 Hz 10 kHz
3.3 to 10.999 nF 0.38 0.01 0.5 0.01 1 pF 50 to 1000 Hz 10 kHz
11 to 32.999 nF 0.19 0.1 0.25 0.1 1 pF 50 to 1000 Hz 10 kHz
33 to 109.99 nF 0.19 0.1 0.25 0.1 10 pF 50 to 1000 Hz 10 kHz
110 to 329.99 nF 0.19 0.3 0.25 0.3 10 pF 50 to 1000 Hz 10 kHz
0.33 to 1.0999 µF 0.19 1 0.25 1 100 pF 50 to 1000 Hz 5 kHz
1.1 to 3.2999 µF 0.26 3 0.35 3 100 pF 50 to 1000 Hz 2 kHz
3.3 to 10.999 µF 0.26 10 0.35 10 1 nF 50 to 400 Hz 1.5 kHz
11 to 32.999 µF 0.30 30 0.40 30 1 nF 50 to 400 Hz 800 Hz
33 to 109.99 µF 0.38 100 0.50 100 10 nF 50 to 200 Hz 400 Hz
110 to 329.99 µF 0.50 300 0.70 300 10 nF 50 to 100 Hz 200 Hz
330 to 1.1 mF 1 300 1 300 100 nF 50 to 100 Hz 150 Hz
Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters.
The output is continuously variable from 330 pF to 1.1 mF.
For all ranges, the maximum charge and discharge current is 150 mA pk or 30 mA rms. The peak voltage
is 4 V, except the 330 µF to 1.1 mF range is limited to 1 V. The maximum lead resistance for no additional
error in 2-wire COMP mode is 10.