EasyManua.ls Logo

Fluke 5500A - Capacitance Accuracy Test

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Service Manual
3-30
Table 3-21. Capacitance Accuracy Test
Nominal Value LCR Stimulus
Frequency
Measured
Value (F)
(NORMAL)
Deviation
%
90-Day Spec.
(%)
0.35 ηF 1 kHz 3.23%
0.48 ηF 1 kHz 2.46
0.6 ηF 1 kHz 2.05
1 ηF 1 kHz 1.38
1.2 ηF 1 kHz 1.22
3 ηF 1 kHz 0.71
3.3 ηF 1 kHz 0.68
10.9 ηF 1 kHz 0.47
12 ηF 1 kHz 1.03
30 ηF 1 kHz 0.52
33 ηF 1 kHz 0.49
109 ηF 1 kHz 0.28
120 ηF 1 kHz 0.44
300 ηF 1 kHz 0.29
330 ηF 100 Hz 0.49
1.09 µF 100 Hz 0.28
1.2 µF 100 Hz 0.51
3 µF 100 Hz 0.36
3.3 µF 100 Hz 0.56
10.9 µF 100 Hz 0.35
12 µF 100 Hz 0.55
30 µF 100 Hz 0.40
33 µF 100 Hz 0.68
109 µF 100 Hz 0.47
120 µF 100 Hz 0.75
300 µF 100 Hz 0.60
330 µF 50 Hz 1.09
1.1 mF 50 Hz 1.03

Table of Contents