SC300 Option
Verification
6
6-99
3. Set HP 8590A start frequency to the Calibrator Mainframe output frequency. Set HP
8590A stop frequency to 10 times the Calibrator Mainframe output frequency. Set
the HP 8590A reference level at +19 dBm.
4. Record the harmonic level reading for each frequency and harmonic listed in Table
6-60. For harmonics 3, 4, and 5, record the highest harmonic level of the three
measured. Harmonics should be below the levels listed in the tolerance column of
Table 6-60.
Table 6-60. Leveled Sine Wave Harmonics Verification
Calibrator Mainframe
Output Frequency
(@ 5.5 V p-p)
Harmonic HP 8590A Reading (dB) Tolerance
50 kHz 2 -33 dB
50 kHz 3, 4, 5 -38 dB
100 kHz 2 -33 dB
100 kHz 3, 4, 5 -38 dB
200 kHz 2 -33 dB
200 kHz 3, 4, 5 -38 dB
400 kHz 2 -33 dB
400 kHz 3, 4, 5 -38 dB
800 kHz 2 -33 dB
800 kHz 3, 4, 5 -38 dB
1 MHz 2 -33 dB
1 MHz 3, 4, 5 -38 dB
2 MHz 2 -33 dB
2 MHz 3, 4, 5 -38 dB
4 MHz 2 -33 dB
4 MHz 3, 4, 5 -38 dB
8 MHz 2 -33 dB
8 MHz 3, 4, 5 -38 dB
10 MHz 2 -33 dB
10 MHz 3, 4, 5 -38 dB
20 MHz 2 -33 dB
20 MHz 3, 4, 5 -38 dB
40 MHz 2 -33 dB
40 MHz 3, 4, 5 -38 dB
80 MHz 2 -33 dB
80 MHz 3, 4, 5 -38 dB
100 MHz 2 -33 dB
100 MHz 3, 4, 5 -38 dB
200 MHz 2 -33 dB
200 MHz 3, 4, 5 -38 dB
250 MHz 2 -33 dB
250 MHz 3, 4, 5 -38 dB