EasyManua.ls Logo

Fluke 5500A - DC Voltage Verification at 1 MΩ

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC300 Option
Verification
6
6-85
Make sure the Calibrator Mainframe impedance is set to 50 (The blue softkey under
Output Z toggles the impedance between 50 and 1 M).
1. Set the HP 3458A to DCV, Auto Range, NPLC = 10, FIXEDZ = on.
2. Program the Calibrator Mainframe to output the voltage listed in Table 6-50. Press
O on the Calibrator Mainframe to activate the output.
3. Allow the HP 3458A reading to stabilize, then record the HP 3458A reading for each
voltage in Table 6-50.
Multiply the readings by (0.5 * (50 + Rload) / Rload), where Rload = the actual
feedthrough termination resistance, to correct for the resistance error. Compare result to
the tolerance (1-year spec.) column.
Table 6-49. DC Voltage Verification at 1 M
Nominal Value (dc) Measured Value (dc) Deviation (mV) 1-Year Spec.
(mV)
0.0 mV 0.10
5.0 mV 0.11
-5.0 mV 0.11
22.0 mV 0.15
-22.0 mV 0.15
25.0 mV 0.16
-25.0 mV 0.16
45.0 mV 0.21
-45.0 mV 0.21
50.0 mV 0.23
-50.0 mV 0.23
220.0 mV 0.65
-220.0 mV 0.65
250.0 mV 0.72
-250.0 mV 0.72
450.0 mV 1.22
-450.0 mV 1.22
500.0 mV 1.35
-500.0 mV 1.35
3.3 V 8.35
-3.3 V 8.35
4.0 V 10.10
-4.0 V 10.10
33.0 V 82.60
-33.0 V 82.60

Table of Contents

Other manuals for Fluke 5500A

Related product manuals