EasyManua.ls Logo

Fluke 5500A - AC Current (Sinewave) Specifications

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Specifications
1
1-13
1-9. AC Current (Sinewave) Specifications
Ranges Frequency
Absolute Uncertainty, tcal ± 5 °C
± (% of output + µA)
Res-
olution
Com-
pliance
Voltage
Max
Induc-
tive
90 days 1 year
Load
0.029 to 0.32999 mA 10 to 20 Hz 0.19% 0.15 µA 0.25% 0.15 µA 0.01 µA 3.0 V rms 1 µH
20 to 45 Hz 0.09 0.15 0.12
5
0.15
45 Hz to 1 kHz 0.09 0.25 0.12
5
0.25
1 to 5 kHz 0.30 0.15 0.4 0.15
5 to 10 kHz 0.94 0.15 1.25 0.15
0.33 to 3.2999 mA 10 to 20 Hz 0.15 0.3 0.2 0.3 0.01 3.0 V rms 1 µH
20 to 45 Hz 0.08 0.3 0.1 0.3
45 Hz to 1 kHz 0.08 0.3 0.1 0.3
1 to 5 kHz 0.15 0.3 0.2 0.3
5 to 10 kHz 0.45 0.3 0.6 0.3
3.3 to 32.999 mA 10 to 20 Hz 0.15 3 0.2 3 0.1 3.0 V rms 200 µH,
20 to 45 Hz 0.08 3 0.1 3 10 to
500 Hz
45 Hz to 1 kHz 0.07 3 0.09 3
1 to 5 kHz 0.15 3 0.2 3 1 µH,
5 to 10 kHz 0.45 3 0.6 3 500 Hz to
10 kHz
33 to 329.99 mA 10 to 20 Hz 0.15 30 0.2 30 1 3.0 to 2.0 V 200 µH,
20 to 45 Hz 0.08 30 0.1 30 rms 10 to
500 Hz
45 Hz to 1 kHz 0.07 30 0.09 30 [1]
1 to 5 kHz 0.15 30 0.2 30 5 µH,
5 to 10 kHz 0.45 30 0.6 30 500 Hz to
10 kHz
0.33 to 2.19999A 10 to 45 Hz 0.15 300 0.2 300 10 3.0 to 2.0 V 200 µH,
45 Hz to 1 kHz
0.08 300 0.1 300 rms
45 to
500 Hz
1 to 5 kHz 0.7 300 0.75 300 [2] 5 µH,
500 Hz to 5
kHz
2.2 to 11A 45 to 65 Hz 0.05 2000 0.06 2000 100 2.8 to 1.25 V 200 µH,
65 to 500 Hz 0.08 2000 0.10 2000 rms 45 to 65 Hz
500 Hz to 1 kHz 0.25 2000 0.33 2000 [3] 1 µH,
65 Hz to 1
kHz