EasyManua.ls Logo

Fluke 5500A - Capacitance Calibration Steps

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration and Verification
Calibration
3
3-13
Table 3-9. Capacitance Calibration Steps
Step 5500A Output (NORMAL) Recommended Stimulus
1 330 ρF 2 V rms at 1 kHz
2 499 ρF
30.5 ηF“
4 1.00 ηF“
5 1.10 ηF“
63.2 ηF“
73.5 ηF“
8 10 ηF“
9 11 ηF“
10 32 ηF“
11 35 ηF“
12 100 ηF“
13 110 ηF“
14 320 ηF 1 V rms at 1 kHz
15 0.35 µF 1 V rms at 100 Hz
16 1 µF“
17 1.1 µF“
18 3.2 µF“
19 3.5 µF“
20 10 µF“
21 11 µF“
22 32 µF“
23 35 µF“
24 100 µF“
25 110 µF“
26 320 µF“
27 350 µF 1 V rms at 50 Hz
28 600 µF“