5500A
Service Manual
3-4
Table 3-1. Required Equipment for Calibration and Verification
Equipment Recommended Model Purpose
Test Lead Kit Fluke 5500A/Leads Provides test cables, esp. TC
leads
8-1/2 digit DMM HP 3458A DC volts, resistance
Mercury Thermometer ASTM 56C Temperature reference
100 mV dc source Fluke 5500A, 5700A, 5440B, or
5100B
Source for thermocouple
measurements (characterize w/
the DMM, if necessary)
Phase Meter Clarke-Hess 6000 Phase
LCR Meter Fluke PM6304C with PM9540/BAN
test lead set
Capacitance
Counter/Timer Fluke PM6666 Frequency
AC Measurement Standard Fluke 5790A ACV and ACI w/ shunts
Shunt Fluke Y5020 10 A dc
Resistance Standard Fluke 742A-1 300 mA dc
Resistance Standard Fluke 742A-10 30 mA dc
Resistance Standard Fluke 742A-100 3 mA dc
Resistance Standard Fluke 742A-10M Resistance at 320 M
Current Shunt Adapter Fluke 792A-7004 Assures compatibility w/ A40
shunts
AC Shunts Fluke A40 (10 mA, 30 mA, 300 mA,
3 A) and A40A-10
ACI
Interconnect cable for A40A Fluke A45-4004 Cable adapter for A40A
Precision metal film
resistors
1 k, 1%, 100 ppm/°C or better Current shunt for <330 µA
(Determine value w/ the DMM)
3-4. Starting Calibration
From the front panel, you start calibration by pressing the S key, followed by the
CAL softkey twice, then 5500A CAL. The CALIBRATION SWITCH on the 5500A rear
panel can be in either position when you begin calibration. It must be set for ENABLE to
store the correction factors into nonvolatile memory.
3-5. How the Calibration Procedure Works
The calibration procedure is self-prompting, with a chance to ABORT and DISCARD
any changes after each function is calibrated. After you press the 5500A CAL softkey,
the procedure works as follows:
1. The 5500A automatically programs the outputs listed in the following tables and
prompts you to make external connections to appropriate measurement instruments.
2. The 5500A then goes into Operate, or asks you to place it into Operate.

Table of Contents