3-47. AC Voltage Harmonic Amplitude Accuracy (NORMAL)
The AC Voltage Harmonic Amplitude Accuracy (NORMAL) tests the accuracy of the
harmonics from the NORMAL terminals. For this test, set the 5500A output to sinewave.
Table 3-34 shows the test points.
Table 3-34. AC Voltage Harmonic Amplitude Accuracy (NORMAL)
Nominal
Value
(NORMAL)
Nominal
Value
(AUX)
Frequency
(AUX)
Harmonic
(NORMAL)
Frequency
(NORMAL)
Measured
Value (V)
(NORMAL)
Deviat
ion %
90-Day
Spec.
(%)
30 mV 300 mV 20 Hz 50th 1 kHz 0.243%
30 mV 300 mV 100 Hz 50th 5 kHz 0.243
30 mV 300 mV 200 Hz 50th 10 kHz 0.243
300 mV 300 mV 20 Hz 50th 1 kHz 0.053
300 mV 300 mV 100 Hz 50th 5 kHz 0.053
300 mV 300 mV 200 Hz 50th 10 kHz 0.053
3 V 3 V 20 Hz 50th 1 kHz 0.024
3 V 3 V 100 Hz 50th 5 kHz 0.024
3 V 3 V 200 Hz 50th 10 kHz 0.024
30 V 3 V 20 Hz 50th 1 kHz 0.034
30 V 3 V 100 Hz 50th 5 kHz 0.034
30 V 3 V 200 Hz 50th 10 kHz 0.034
300 V 3 V 50 Hz 20th 1 kHz 0.044
300 V 3 V 100 Hz 50th 5 kHz 0.070
300 V 3 V 200 Hz 50th 10 kHz 0.070
1000 V 3 V 50 Hz 20th 1 kHz 0.056
1000 V 3 V 100 Hz 50th 5 kHz 0.170
800 V 3 V 200 Hz 50th 10 kHz 0.275
Optional:
1000 V
3 V 200 Hz 50th 10 kHz 0.250