Introduction and Specifications
Specifications
1
1-19
1-13. DC Power Specification Summary
5500A Calibrator Current Range
Voltage Range 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
90 days 33 mV to 1020 V 0.03% 0.02% 0.03% 0.02%
1 year 33 mV to 1020 V 0.04% 0.03% 0.04% 0.03%
5500A Calibrator Current Range
Voltage Range 0.33 to 0.8999 A 0.9 to 2.1999 A 2.2 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
90 days 33 mV to 1020 V 0.07% 0.05% 0.08% 0.06%
1 year 33 mV to 1020 V 0.08% 0.06% 0.12% 0.09%
5725A Amplifier Current Range
Voltage Range 1.5 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
90 days 33 mV to 1020 V 0.09% 0.07%
1 year 33 mV to 1020 V 0.10% 0.08%
Note
1. To determine dc power uncertainty with more precision, see the individual “DC
Voltage Specifications” and “DC Current Specifications” and “Calculating Power
Uncertainty.”