Fluke 5500A - Time Marker Specifications

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC600 Option
SC600 Specifications
6
6-9
6-7. Time Marker Specifications
Table 6-4. Time Marker Specifications
Time Marker
into 50
5 s to 50 ms 20 ms to
100 ns
50 ns to
20 ns
10 ns 5 ns to
2 ns
1-Year Absolute
Uncertainty at Cardinal
Points, tcal ±5 °C [3]
±(25 + t * 1000)
ppm [1]
± 2.5 ppm ± 2.5 ppm ± 2.5 ppm ± 2.5 ppm
Wave Shape spike or square spike, square,
or 20%-pulse
spike or
square
square or
sine
sine
Typical Output Level > 1 V p-p [2] > 1 V p-p [2] > 1 V p-p [2] >1 V p-p [2] > 1 V p-p
Typical Jitter (rms) <10 ppm < 1 ppm < 1 ppm <1 ppm <1 ppm
Sequence (cardinal
points)
5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
Adjustment Range At least ± 10% around each cardinal points.
Amplitude Resolution 4 digits
[1] t is time in seconds. Examples: At 5 s the uncertainty is 5,025 ppm; At 50 ms the uncertainty is 75 ppm.
[2] Typical rise time of square wave and 20%-pulse (20% duty cycle pulse) is < 1.5 ns.
[3] Away from the cardinal points, add ±50 ppm to uncertainty
.
6-8. Wave Generator Specifications
Table 6-5. Wave Generator Specifications
Wave Generator Characteristics
Square Wave, Sine Wave, and Triangle Wave
into 50 or 1 M
Amplitude
Range into 1 M: 1.8 mV to 55 V p-p
into 50 : 1.8 mV to 2.5 V p-p
1-Year Absolute Uncertainty, tcal ±
5 °C,
10 Hz to 10 kHz
± (3% of p-p output + 100 µV)
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical DC Offset Range 0 to ± (40% of p-p amplitude) [1]
Frequency
Range 10 Hz to 100 kHz
Resolution 4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal ± 5 °C ± (25 ppm + 15 mHz)
[1]The DC offset plus the wave signal must not exceed 30 V rms.